105
19.
Open the lid of the
luminometer.
Remove the test tube from
the luminometer and put it
back in the LUMIStherm.
20.
Do steps
15
and
19
again until:
•
All the test tubes in Row
B and C have been
measured and recorded
moving from left to right
(e.g., B2, C2, B3, C3,
etc.).
•
If testing D 1 values and
higher, 0.8 mL of the
sample dilution from the
test tube in position A2
has been added to the
test tubes in position B2
and C2.
•
If testing D 2 values and
higher, 0.5 mL of the
sample dilution from the
test tube in position A2
has been added to the
test tubes in position B2
and C2.
•
0.5 mL of the sample
dilution from each test
tube in Row A (A3 and
higher) has been added
to the test tubes in Row
B and C that have the
same position number
(e.g., from A
4
to B
4
and
C
4
).
Note:
Add the sample dilution
to each test tube in Row B and
Row C immediately after the
test tube is measured.