50
9.3 TDS/Conductivity calibration
7.
Wait for the value to
stabilize, then push the
Calibration/Settings key to
save the buffer value.
8.
Optional: Do steps 2
through 7 againto measure
a pH 4.00 and/or a pH
10.00 buffer.
9.
Push and hold the
Calibration/Settings key
to exit.
1.
Push the
Power/Backlight key to set
the power to on. Remove
the sensor cap.
2.
Rinse the sensor and the
sensor cap with deionized
water. Dry with a no-lint
cloth.
3.
Push and hold the
Lock/Parameter key until
“Cond” shows as the
measurement mode.