103
7.
While B1 is being
measured, set the timer to
the correct contact time
(e.g., 15 or 30 minutes).
Start the timer.
When the measurement is
complete (approximately
15 seconds), the
luminometer shows the
relative light intensity of
the test tube.
8.
Record the measured
value on a sheet of paper if
the measured value is
shown but not saved to the
luminometer, saved to the
computer or printed.
9.
Open the lid of the
luminometer.
Remove the test tube from
the luminometer and put it
back in the LUMIStherm.
10.
Put the test tube in
position C1 into the black
test tube holder in the
luminometer cell.
11.
Close the luminometer
lid. Push
MEASURE
.
12.
If testing D 2 values
and higher, put the pipette
in test tube A1 and pull in
0.5 mL of sample dilution.
If testing D 1 values and
higher, put the pipette in
test tube A1 and pull in 0.8
mL of sample dilution.
Put the pipette into the test
suspension and slowly
dispense the sample in B1.
Mix with the pipette.