21
To show or change the waiting time and measuring time:
1.
Push
ON
(green button) to apply power to the instrument.
The instrument does built-in tests that make sure that the electronics and
software are operating correctly.
2.
When the built-in tests are done, push
PROCEED
.
The Main Menu is shown.
3.
Select Luminescent Bacteria Test and push
ENTER
.
4.
Select Set-Up and push
ENTER
.
The Set-up Menu is shown.
5.
To show or change the waiting time:
a.
Select Set Waiting Time and push
ENTER
.
The current settings are shown.
b.
To change the waiting time, push
CHANGE
.
c.
Push
STORE
to save the change.
The Set-up Menu is shown.
6.
To show or change the measuring time:
a.
Select Set Measuring Time and push
ENTER
.
The current settings are shown.
b.
To change the measuring time, push
CHANGE
.
c.
Push
STORE
to save the change.
The Set-up Menu is shown.
3.4 Connect the luminometer to a printer
Luminometer measurements can be sent to a printer either during the test or after
the test is done.
To connect the luminometer to a printer:
1.
Pull out the plug that is attached to the lanyard.
2.
If using a DPU-414 thermal printer, turn off the printer.
3.
Connect the RS232 serial interface cable to the luminometer.
Summary of Contents for Eclox
Page 1: ...2886888 Eclox Rapid Response Test Kit User Manual 01 2018 Edition 5 ...
Page 2: ...Visit us at www hach com ...
Page 6: ...Table of Contents Table of Contents Page 6 EcloxTOC fm ...
Page 14: ...14 ...
Page 24: ...24 ...
Page 46: ...46 ...
Page 52: ...52 ...
Page 110: ...110 ...
Page 113: ...113 ...
Page 114: ...Visit us at www hach com ...
Page 120: ...120 ...
Page 123: ...123 ...