Estimated Detection Limit
The estimated detection limit for program #1 is 0.013 mg/L Al. For more
information on the estimated detection limit, see
Section 1
.
Interferences
Fluoride interferes at all levels by complexing with aluminum. The actual
aluminum concentration can be determined using the Fluoride
Interference Graph when the fluoride concentration is known. To use the
fluoride interference graph:
Interfering
Substance
Interference Levels and Treatments
Acidity
Acidity interferes at greater than 300 mg/L as CaCO
3
. Treat sam-
ples with greater than 300 mg/L acidity as CaCO
3
as follows:
1.
Add one drop of m-Nitrophenol Indicator Solution to the sample
taken in Step 3.
2.
Add one drop of 5.0 N Sodium Hydroxide Standard Solution.
Stopper the cylinder. Invert to mix. Repeat as often as necessary
until the color changes from colorless
to yellow.
3.
Add one drop of 5.25 N Sulfuric Acid Standard Solution to
change the solution from yellow back to colorless. Continue with
the test.
Alkalinity
1000 mg/L as CaCO
3
. Eliminate interferences from higher alkalin-
ity concentrations using the following pretreatment:
1.
Add one drop of m-Nitrophenol Indicator Solution to the sample
taken in Step 3. A yellow color indicates
excessive alkalinity.
2.
Add one drop of 5.25 N Sulfuric Acid Standard Solution. Stop-
per the cylinder. Invert to mix. If the yellow color persists, repeat
until the sample becomes colorless. Continue with the test.
Calcium
Does not interfere.
Fluoride
Interferes at all levels. See graph below.
Iron
Greater than 20 mg/L.
Phosphate
Greater than 50 mg/L.
Polyphosphate
Polyphosphate interferes at all levels by causing negative errors and
must not be present. Before running the test, polyphosphate must
be converted to orthophosphate by acid hydrolysis as described
under the phosphorus procedures.
ALUMINUM,
continued
Summary of Contents for DR/850
Page 2: ...2...
Page 8: ...8...
Page 9: ...9 Sample Procedure Explained...
Page 10: ...10 Sample Procedure Explained continued...
Page 11: ...11 Sample Procedure Explained continued...
Page 12: ...12 Sample Procedure Explained continued...
Page 41: ...41 CHEMICAL ANALYSIS INFORMATION continued Figure 10 Multiple Standard Additions Graph...
Page 69: ...69 SECTION 4 PROCEDURES...
Page 70: ...70...
Page 92: ...92...
Page 100: ...100...
Page 108: ...108...
Page 116: ...116...
Page 139: ...139...
Page 140: ...140...
Page 168: ...168...
Page 176: ...176...
Page 184: ...184...
Page 194: ...194...
Page 202: ...202...
Page 214: ...214...
Page 234: ...234...
Page 242: ...242...
Page 264: ...264...
Page 270: ...270...
Page 284: ...284...
Page 298: ...298...
Page 314: ...314...
Page 324: ...324...
Page 332: ...332...
Page 336: ...336...
Page 340: ...340...
Page 354: ...354...
Page 370: ...370...
Page 378: ...378...
Page 386: ...386...
Page 398: ...398...
Page 404: ...404...
Page 412: ...412...
Page 424: ...424...
Page 436: ...436...
Page 446: ...446...
Page 456: ......
Page 459: ......
Page 460: ...460...