197
5.
Press:
ZERO
The cursor will move to
the right, then the
display will show:
0.00 mg/L Fe
6.
Fill a FerroVer
AccuVac Ampul
with sample.
Note:
Keep the tip
immersed while the ampul
fills completely.
7.
Quickly invert the
ampul several times to
mix. Wipe off any liquid
or fingerprints.
Note:
An orange color will
form if iron is present.
Note:
Accuracy is not
affected by undissolved
powder.
8.
Press:
TIMER ENTER
A three-minute reaction
period will begin.
Note:
Samples containing
visible rust should be
allowed to react at least
five minutes.
9.
Place the AccuVac
ampul into the cell
holder. Tightly cover the
ampul with the
instrument cap.
10.
Press:
READ
The cursor will move to
the right, then the result
in mg/L iron (Fe) will
be displayed.
Note:
Standard Adjust may
be performed using a
prepared standard (see
Section 1).
IRON, TOTAL,
continued
Summary of Contents for DR/850
Page 2: ...2...
Page 8: ...8...
Page 9: ...9 Sample Procedure Explained...
Page 10: ...10 Sample Procedure Explained continued...
Page 11: ...11 Sample Procedure Explained continued...
Page 12: ...12 Sample Procedure Explained continued...
Page 41: ...41 CHEMICAL ANALYSIS INFORMATION continued Figure 10 Multiple Standard Additions Graph...
Page 69: ...69 SECTION 4 PROCEDURES...
Page 70: ...70...
Page 92: ...92...
Page 100: ...100...
Page 108: ...108...
Page 116: ...116...
Page 139: ...139...
Page 140: ...140...
Page 168: ...168...
Page 176: ...176...
Page 184: ...184...
Page 194: ...194...
Page 202: ...202...
Page 214: ...214...
Page 234: ...234...
Page 242: ...242...
Page 264: ...264...
Page 270: ...270...
Page 284: ...284...
Page 298: ...298...
Page 314: ...314...
Page 324: ...324...
Page 332: ...332...
Page 336: ...336...
Page 340: ...340...
Page 354: ...354...
Page 370: ...370...
Page 378: ...378...
Page 386: ...386...
Page 398: ...398...
Page 404: ...404...
Page 412: ...412...
Page 424: ...424...
Page 436: ...436...
Page 446: ...446...
Page 456: ......
Page 459: ......
Page 460: ...460...