44
CHEMICAL ANALYSIS INFORMATION
, continued
Hach provides a value called the Estimated Detection Limit (EDL) for all
programs. It is the calculated lowest average concentration in a deionized
water matrix that is different from zero with a 99% level of confidence.
Specifically, it is the upper 99% confidence limit for zero concentration
based on the calibration data used to prepare the pre-programmed
calibration curve.
Do not use the EDL as the MDL
. The conditions for
MDL determination must be exactly the same as the conditions used for
analysis. The EDL may be useful to the analyst as a starting point in
determining a MDL or as a way to compare methods. Measurements
below the EDL may also be valuable because they can show a trend,
indicate the presence of analyte and/or provide statistical data. However,
these values have a large uncertainty.
Method Detection Limit (MDL)
This method is in accordance with the USEPA definition in 40 CFR, Part
136, Appendix B (see most current edition).
The USEPA defines the method detection limit (MDL) as the minimum
concentration that can be determined with 99% confidence that the true
concentration is greater than zero. Since the MDL will vary from analyst
to analyst, it is important that analysts determine the MDL based on their
unique operating conditions.
The procedure for determining MDL is based on replicate analyses at a
concentration 1 to 5 times the estimated detection limit. The MDL value
is calculated from the standard deviation of the replicate study results
multiplied by the appropriate Student’s
t
value for a 99% confidence
interval. For this definition, the MDL does not account for variation in
sample composition and can only be achieved under ideal conditions.
1.
Estimate the detection limit. Use the Hach estimated detection
limit (EDL) value stated in the
Method Performance
section of the
analysis procedure.
2.
Prepare a laboratory standard of the analyte in deionized water which
is free of the analyte that is 1 to 5 times the estimated detection limit.
3.
Analyze at least seven portions of the laboratory standard and record
each result.
4.
Calculate the average and standard deviation
(s)
of the results.
Summary of Contents for DR/850
Page 2: ...2...
Page 8: ...8...
Page 9: ...9 Sample Procedure Explained...
Page 10: ...10 Sample Procedure Explained continued...
Page 11: ...11 Sample Procedure Explained continued...
Page 12: ...12 Sample Procedure Explained continued...
Page 41: ...41 CHEMICAL ANALYSIS INFORMATION continued Figure 10 Multiple Standard Additions Graph...
Page 69: ...69 SECTION 4 PROCEDURES...
Page 70: ...70...
Page 92: ...92...
Page 100: ...100...
Page 108: ...108...
Page 116: ...116...
Page 139: ...139...
Page 140: ...140...
Page 168: ...168...
Page 176: ...176...
Page 184: ...184...
Page 194: ...194...
Page 202: ...202...
Page 214: ...214...
Page 234: ...234...
Page 242: ...242...
Page 264: ...264...
Page 270: ...270...
Page 284: ...284...
Page 298: ...298...
Page 314: ...314...
Page 324: ...324...
Page 332: ...332...
Page 336: ...336...
Page 340: ...340...
Page 354: ...354...
Page 370: ...370...
Page 378: ...378...
Page 386: ...386...
Page 398: ...398...
Page 404: ...404...
Page 412: ...412...
Page 424: ...424...
Page 436: ...436...
Page 446: ...446...
Page 456: ......
Page 459: ......
Page 460: ...460...