200
Summary of Method
FerroVer Iron Reagent reacts with all soluble iron and most
insoluble forms of iron in the sample to produce soluble ferrous
iron. This reacts with 1,10-phenanthroline indicator in the reagent
to form an orange color in proportion to the iron concentration.
REQUIRED REAGENTS & APPARATUS (Using Powder Pillows)
Quantity Required
Description
Per Test
Unit
Cat No.
FerroVer Iron Reagent Powder Pillows ...................1 pillow ............... 100/pkg.......... 21057-69
Sample cell, 10-20-25 mL, with screw cap .................. 1......................... 6/pkg.......... 24019-06
REQUIRED REAGENTS & APPARATUS (Using AccuVac Ampuls)
FerroVer Iron Reagent AccuVac Ampuls ................1 ampul ................. 25/pkg.......... 25070-25
Beaker, 50 mL............................................................... 1........................... each........... 500-41H
OPTIONAL REAGENTS
Description
Unit
Cat. No.
Ammonium Hydroxide, ACS ................................................................500 mL.............. 106-49
Drinking Water Standard, Metals, LR (Cu, Fe, Mn) .............................500 mL.......... 28337-49
Drinking Water Standard, Metals, HR (Cu, Fe, Mn) .............................500 mL.......... 28336-49
Hydrochloric Acid Standard Solution, 6 N............................................500 mL.............. 884-49
Hydrochloric Acid, ACS........................................................................500 mL.............. 134-49
Iron Standard Solution, 100 mg/L .........................................................100 mL.......... 14175-42
Iron Ampule Standard, 50 mg/L ............................................................ 20/pkg.......... 14254-20
Nitric Acid, ACS....................................................................................500 mL.............. 152-49
Nitric Acid Solution, 1:1........................................................................500 mL............ 2540-49
RoVer Rust Remover ................................................................................ 454 g.............. 300-01
Sodium Hydroxide Standard Solution, 5.0 N .............................100 mL MDB............ 2450-32
Water, deionized............................................................................................ 4 L.............. 272-56
Turbidity
1.
Add 0.1 g scoop of RoVer Rust Remover to the blank
in Step 3. Swirl to mix.
2.
Zero the instrument with this blank.
3.
If sample remains turbid, add three 0.2 g scoops of
RoVer to a 75-mL sample. Let stand
5 minutes.
4.
Filter through a glass filter or centrifuge.
5.
Use filtered sample in Steps 3 and 6.
Sample pH (extreme)
Adjust pH to 3-5. See
Interferences
in
Section 1.
Highly Buffered Samples
Adjust pH to 3-5. See
Interferences
in
Section 1.
Interfering Substance
Interference Level and Treatment
IRON, TOTAL,
continued
Summary of Contents for DR/850
Page 2: ...2...
Page 8: ...8...
Page 9: ...9 Sample Procedure Explained...
Page 10: ...10 Sample Procedure Explained continued...
Page 11: ...11 Sample Procedure Explained continued...
Page 12: ...12 Sample Procedure Explained continued...
Page 41: ...41 CHEMICAL ANALYSIS INFORMATION continued Figure 10 Multiple Standard Additions Graph...
Page 69: ...69 SECTION 4 PROCEDURES...
Page 70: ...70...
Page 92: ...92...
Page 100: ...100...
Page 108: ...108...
Page 116: ...116...
Page 139: ...139...
Page 140: ...140...
Page 168: ...168...
Page 176: ...176...
Page 184: ...184...
Page 194: ...194...
Page 202: ...202...
Page 214: ...214...
Page 234: ...234...
Page 242: ...242...
Page 264: ...264...
Page 270: ...270...
Page 284: ...284...
Page 298: ...298...
Page 314: ...314...
Page 324: ...324...
Page 332: ...332...
Page 336: ...336...
Page 340: ...340...
Page 354: ...354...
Page 370: ...370...
Page 378: ...378...
Page 386: ...386...
Page 398: ...398...
Page 404: ...404...
Page 412: ...412...
Page 424: ...424...
Page 436: ...436...
Page 446: ...446...
Page 456: ......
Page 459: ......
Page 460: ...460...