337
Method 10076
pH
(6.5 to 8.5 pH units)
Colorimetric pH Determination Using Phenol Red
For water and wastewater
1.
Enter the stored
program number for the
pH method.
Press:
PRGM
The display will show:
PRGM
?
2.
Press:
75 ENTER
The display will show
PH
and the
ZERO
icon.
3.
Fill a sample cell
with 10 mL of sample
(the blank).
4.
Place the blank in the
cell holder. Tightly
cover the sample cell
with the instrument cap.
5.
Press:
ZERO
The cursor will move to
the right, then the
display will show:
6.0 PH
6.
Fill another cell with
10 mL of sample.
Note:
Sample temperature
must be 21-29 °C.
7.
Using a disposable
dropper, add 1 mL of
Phenol Red Indicator
Solution to the cell (the
prepared sample). Cap
the sample cell and
invert twice to mix.
8.
Place the prepared
sample into the cell
holder. Tightly cover the
sample cell with the
instrument cap.
Summary of Contents for DR/850
Page 2: ...2...
Page 8: ...8...
Page 9: ...9 Sample Procedure Explained...
Page 10: ...10 Sample Procedure Explained continued...
Page 11: ...11 Sample Procedure Explained continued...
Page 12: ...12 Sample Procedure Explained continued...
Page 41: ...41 CHEMICAL ANALYSIS INFORMATION continued Figure 10 Multiple Standard Additions Graph...
Page 69: ...69 SECTION 4 PROCEDURES...
Page 70: ...70...
Page 92: ...92...
Page 100: ...100...
Page 108: ...108...
Page 116: ...116...
Page 139: ...139...
Page 140: ...140...
Page 168: ...168...
Page 176: ...176...
Page 184: ...184...
Page 194: ...194...
Page 202: ...202...
Page 214: ...214...
Page 234: ...234...
Page 242: ...242...
Page 264: ...264...
Page 270: ...270...
Page 284: ...284...
Page 298: ...298...
Page 314: ...314...
Page 324: ...324...
Page 332: ...332...
Page 336: ...336...
Page 340: ...340...
Page 354: ...354...
Page 370: ...370...
Page 378: ...378...
Page 386: ...386...
Page 398: ...398...
Page 404: ...404...
Page 412: ...412...
Page 424: ...424...
Page 436: ...436...
Page 446: ...446...
Page 456: ......
Page 459: ......
Page 460: ...460...