Sampling and Storage
Collect samples in acid-cleaned glass or plastic containers. No
acid addition is necessary if analyzing the sample immediately.
To preserve samples, adjust the pH to 2 or less with nitric acid
(about 2 mL per liter). Preserved samples may be stored up to six
months at room temperature. Adjust the pH to between 3 and 5
with 5.0 N Sodium Hydroxide Standard Solution before analysis.
Correct the test result for volume additions; see
Correcting for
Volume Additions
in
Section 1
for more information. If only
dissolved iron is to be determined, filter the sample before
adding the acid.
Accuracy Check
Standard Additions Method
a)
Snap the neck off a 50 mg/L Iron PourRite Ampule
Standard Solution.
b)
Use the TenSette Pipet to add 0.1, 0.2, and 0.3 mL of
standard, respectively, to three 25-mL samples and mix
thoroughly.
c)
For analysis using AccuVac Ampuls, transfer solutions to
dry, clean 50-mL beakers to facilitate filling of the ampuls.
For analysis with powder pillows, transfer only 10 mL of
solution to the 10-mL sample cells.
d)
Analyze each standard addition sample as described
above. The iron concentration should increase 0.2 mg/L
for each 0.1 mL of standard added.
e)
If these increases do not occur, see
Standard Additions
in
Section 1
for troubleshooting information.
Standard Solution Method
Prepare a 1.0-mg/L iron standard by diluting 1.00 mL of Iron
Standard Solution, 100 mg/L Fe, to 100 mL with deionized water.
Or, dilute 1.00 mL of an Iron PourRite Ampule Standard Solution
(50 mg/L) to 50 mL in a volumetric flask. Prepare this solution daily.
Run the test following the procedure for powder pillows or
AccuVac Ampuls. Results should be between 0.90 mg/L and 1.10
mg/L Fe.
IRON, TOTAL,
continued
Summary of Contents for DR/850
Page 2: ...2...
Page 8: ...8...
Page 9: ...9 Sample Procedure Explained...
Page 10: ...10 Sample Procedure Explained continued...
Page 11: ...11 Sample Procedure Explained continued...
Page 12: ...12 Sample Procedure Explained continued...
Page 41: ...41 CHEMICAL ANALYSIS INFORMATION continued Figure 10 Multiple Standard Additions Graph...
Page 69: ...69 SECTION 4 PROCEDURES...
Page 70: ...70...
Page 92: ...92...
Page 100: ...100...
Page 108: ...108...
Page 116: ...116...
Page 139: ...139...
Page 140: ...140...
Page 168: ...168...
Page 176: ...176...
Page 184: ...184...
Page 194: ...194...
Page 202: ...202...
Page 214: ...214...
Page 234: ...234...
Page 242: ...242...
Page 264: ...264...
Page 270: ...270...
Page 284: ...284...
Page 298: ...298...
Page 314: ...314...
Page 324: ...324...
Page 332: ...332...
Page 336: ...336...
Page 340: ...340...
Page 354: ...354...
Page 370: ...370...
Page 378: ...378...
Page 386: ...386...
Page 398: ...398...
Page 404: ...404...
Page 412: ...412...
Page 424: ...424...
Page 436: ...436...
Page 446: ...446...
Page 456: ......
Page 459: ......
Page 460: ...460...