275
Summary of Method
Ammonia compounds combine with chlorine to form
monochloramine. Monochloramine reacts with salicylate to form 5-
aminosalicylate. The 5-aminosalicylate is oxidized in the presence of a
sodium nitroprusside catalyst to form a blue-colored compound. The
blue color is masked by the yellow color from the excess reagent
present to give a green-colored solution.
Safety
Good safety habits and laboratory techniques should be used
throughout the procedure. Consult the
Material Safety Data
Sheets
for information specific to the reagents used. For
additional information, refer to
Section 3
.
Pollution Prevention And Waste Management
The ammonia salicylate reagent contains sodium
nitroferricyanide. Cyanide solutions are regulated as hazardous
wastes by the Federal RCRA. Collect cyanide solutions for
disposal as reactive (D001) waste. Be sure cyanide solutions are
stored in a caustic solution with pH >11 to prevent release of
hydrogen cyanide gas. See
Section 3
for further information in
proper disposal of these materials.
REQUIRED REAGENTS
AmVer
™
Reagent Set for Nitrogen, Ammonia, High Range, TNT (25 tests) ..............26069-45
Includes: (1) 23952-66, (1) 23954-66, (1) 272-42,
*
(50) AmVer HR Vials
Quantity Required
Description
Per Test
Unit
Cat. No.
AmVer
™
HR Reagent Test ‘N Tube
™
Vials ............ 2 vials ...................50/pkg ....................... *
Ammonia Salicylate Reagent Powder Pillows....... 2 pillows.................50/pkg ..........23952-66
Ammonia Cyanurate Reagent Powder Pillows ...... 2 pillows.................50/pkg ..........23954-66
REQUIRED APPARATUS
COD/TNT Adapter........................................................... 1 .................each .................48464-00
Pipet, TenSette
®
, 0-1 mL ................................................. 1 .................each .................19700-01
Pipet Tips for 19700-01................................................ varies ......... 50/pkg .................21856-96
Test Tube Rack ............................................................... 1-3 ...............each .................18641-00
Funnel, micro (for reagent addition) ................................ 1 .................each .................25843-35
* Not available separately.
NITROGEN, AMMONIA, High Range, Test 'N Tube,
continued
Summary of Contents for DR/850
Page 2: ...2...
Page 8: ...8...
Page 9: ...9 Sample Procedure Explained...
Page 10: ...10 Sample Procedure Explained continued...
Page 11: ...11 Sample Procedure Explained continued...
Page 12: ...12 Sample Procedure Explained continued...
Page 41: ...41 CHEMICAL ANALYSIS INFORMATION continued Figure 10 Multiple Standard Additions Graph...
Page 69: ...69 SECTION 4 PROCEDURES...
Page 70: ...70...
Page 92: ...92...
Page 100: ...100...
Page 108: ...108...
Page 116: ...116...
Page 139: ...139...
Page 140: ...140...
Page 168: ...168...
Page 176: ...176...
Page 184: ...184...
Page 194: ...194...
Page 202: ...202...
Page 214: ...214...
Page 234: ...234...
Page 242: ...242...
Page 264: ...264...
Page 270: ...270...
Page 284: ...284...
Page 298: ...298...
Page 314: ...314...
Page 324: ...324...
Page 332: ...332...
Page 336: ...336...
Page 340: ...340...
Page 354: ...354...
Page 370: ...370...
Page 378: ...378...
Page 386: ...386...
Page 398: ...398...
Page 404: ...404...
Page 412: ...412...
Page 424: ...424...
Page 436: ...436...
Page 446: ...446...
Page 456: ......
Page 459: ......
Page 460: ...460...