335
Estimated Detection Limit
The estimated detection limit for the programs #72, #73, and #74
is 0.02 mg/L O
3
. For more information on the estimated detection
limit, see
Section 1
.
Summary of Method
The reagent formulation adjusts the sample pH to 2.5 after the
ampul has filled. The indigo reagent reacts immediately and
quantitatively with ozone. The blue color of indigo is bleached in
proportion to the amount of ozone present in the sample. Other
reagents in the formulation prevent chlorine interference. No
transfer of sample is needed in the procedure. Therefore, ozone
loss due to sampling is eliminated.
REQUIRED REAGENTS
Quantity Required
Description
Per Test
Unit
Cat. No.
Ozone AccuVac Ampuls
Select one or more based on range:
0-0.25 mg/L............................................................ 2 ampuls .................25/pkg ..........25160-25
0-0.75 mg/L............................................................ 2 ampuls .................25/pkg ..........25170-25
0-1.50 mg/L............................................................ 2 ampuls .................25/pkg ..........25180-25
Water, deionized ........................................................ varies ........................ 4 L ..............272-56
REQUIRED APPARATUS
Beaker, 50 mL ...............................................................2 ...........................each ...........500-41H
OPTIONAL APPARATUS
AccuVac Snapper Kit ..................................................................................each ..........24052-00
AccuVac Ampule sampler...........................................................................each ..........24051-00
For Technical Assistance, Price and Ordering
In the U.S.A.—Call 800-227-4224
Outside the U.S.A.—Contact the Hach office or distributor serving you.
OZONE,
continued
Summary of Contents for DR/850
Page 2: ...2...
Page 8: ...8...
Page 9: ...9 Sample Procedure Explained...
Page 10: ...10 Sample Procedure Explained continued...
Page 11: ...11 Sample Procedure Explained continued...
Page 12: ...12 Sample Procedure Explained continued...
Page 41: ...41 CHEMICAL ANALYSIS INFORMATION continued Figure 10 Multiple Standard Additions Graph...
Page 69: ...69 SECTION 4 PROCEDURES...
Page 70: ...70...
Page 92: ...92...
Page 100: ...100...
Page 108: ...108...
Page 116: ...116...
Page 139: ...139...
Page 140: ...140...
Page 168: ...168...
Page 176: ...176...
Page 184: ...184...
Page 194: ...194...
Page 202: ...202...
Page 214: ...214...
Page 234: ...234...
Page 242: ...242...
Page 264: ...264...
Page 270: ...270...
Page 284: ...284...
Page 298: ...298...
Page 314: ...314...
Page 324: ...324...
Page 332: ...332...
Page 336: ...336...
Page 340: ...340...
Page 354: ...354...
Page 370: ...370...
Page 378: ...378...
Page 386: ...386...
Page 398: ...398...
Page 404: ...404...
Page 412: ...412...
Page 424: ...424...
Page 436: ...436...
Page 446: ...446...
Page 456: ......
Page 459: ......
Page 460: ...460...