ORGANIC CARBON, TOTAL, High Range,
continued
If the sample contains greater than 600 mg/L CaCO
3
alkalinity,
lower the sample pH to less than 7 before testing by adding
sulfuric acid solution.
Most sample turbidity is either dissolved during the digestion
stage or settled during the cooling period. Sample turbidities up
to 900 NTU have been tested without interference.
Summary of Method
The total organic carbon (TOC) is determined by first sparging
the sample under slightly acidic conditions to remove the
inorganic carbon. In the outside vial, organic carbon in the sample
is digested by persulfate and acid to form carbon dioxide. During
digestion, the carbon dioxide diffuses into a pH indicator reagent
in the inner ampule. The adsorption of carbon dioxide into the
indicator forms carbonic acid. Carbonic acid changes the pH of
the indicator solution which, in turn, changes the color. The
amount of color change is related to the original amount of
carbon present in the sample.
REQUIRED REAGENTS
Total Organic Carbon Direct Method High Range
Test ’N Tube Reagent Set................................................................50 vials.......... 27604-45
Includes:
Description
Qty/Test
Unit
Cat. No.
Acid Digestion Solution Vials, High Range TOC .................. 1 ............. 50/pkg
............................. *
Buffer Solution, Sulfate ..................................................... 0.4 mL .........25 mL.............. 452-33
Funnel, micro .......................................................................... 1 ................. each.......... 25843-35
Indicator Ampules, High Range TOC .................................... 1 ............. 10/pkg....................... *
TOC Persulfate Powder Pillows ............................................. 1 ............. 50/pkg....................... *
Water, organic-free
**
......................................................... 0.3 mL .......500 mL.......... 26415-49
Silica
100 mg/L SiO
2
Sulfate
5000 mg/L SO
4
2-
Sulfide
20 mg/L S
2-
Sulfite
50 mg/L SO
3
2-
Zinc
5 mg/L
Table 1 Non-interfering Substances (Continued)
Substance
Maximum Level Tested
* These items are not sold separately.
Summary of Contents for DR/850
Page 2: ...2...
Page 8: ...8...
Page 9: ...9 Sample Procedure Explained...
Page 10: ...10 Sample Procedure Explained continued...
Page 11: ...11 Sample Procedure Explained continued...
Page 12: ...12 Sample Procedure Explained continued...
Page 41: ...41 CHEMICAL ANALYSIS INFORMATION continued Figure 10 Multiple Standard Additions Graph...
Page 69: ...69 SECTION 4 PROCEDURES...
Page 70: ...70...
Page 92: ...92...
Page 100: ...100...
Page 108: ...108...
Page 116: ...116...
Page 139: ...139...
Page 140: ...140...
Page 168: ...168...
Page 176: ...176...
Page 184: ...184...
Page 194: ...194...
Page 202: ...202...
Page 214: ...214...
Page 234: ...234...
Page 242: ...242...
Page 264: ...264...
Page 270: ...270...
Page 284: ...284...
Page 298: ...298...
Page 314: ...314...
Page 324: ...324...
Page 332: ...332...
Page 336: ...336...
Page 340: ...340...
Page 354: ...354...
Page 370: ...370...
Page 378: ...378...
Page 386: ...386...
Page 398: ...398...
Page 404: ...404...
Page 412: ...412...
Page 424: ...424...
Page 436: ...436...
Page 446: ...446...
Page 456: ......
Page 459: ......
Page 460: ...460...