181
Interferences
This test is sensitive to small amounts of interference. Glassware
must be very clean. Repeating the test with the same glassware is
recommended to ensure that results are accurate.
The following substances interfere to the extent shown:
SPADNS 2 Reagent contains enough non-toxic reducing agent to
eliminate interference up to 5 mg/L chlorine. For higher chlorine
levels, dilute sample with deionized water by a factor that will
lower chlorine concentration to below 5 mg/L. Perform the
procedure, and multiply results by this factor to obtain
mg/L Fluoride.
To check for interferences from aluminum, read the concentration
one minute after reagent addition, then again after 15 minutes.
An appreciable increase in concentration suggests aluminum
interference. Waiting two hours before making the final reading
will eliminate the effect of up to 3.0 mg/L aluminum.
Most interferences can be eliminated by distilling the sample
from an acid solution as described below:
a)
Set up the distillation apparatus for the general purpose
distillation. See the Hach Distillation Apparatus Manual.
Turn on the water and make certain it is flowing through
the condenser.
b)
Measure 100 mL of sample into the distillation flask.
Add a magnetic stirring bar and turn on the heater power
switch.
Turn the stir control to 5.
c)
Cautiously measure 150 mL of StillVer Distillation
Solution (2:1 Sulfuric Acid) into the flask. If high levels
of chloride are present, add 5 mg silver sulfate for each
Substance
Concentration
Error
Alkalinity (as CaCO
3
)
5000 mg/L
-0.1 mg/L F
-
Aluminum
0.1 mg/L
-0.1 mg/L F
-
Chloride
7000 mg/L
+0.1 mg/L F
-
Iron, ferric
10 mg/L
-0.1 mg/L F
-
Phosphate, ortho
16 mg/L
+0.1 mg/L F
-
Sodium Hexametaphosphate
1.0 mg/L
+0.1 mg/L F
-
Sulfate
200 mg/L
+0.1 mg/L F
-
FLUORIDE,
continued
Summary of Contents for DR/850
Page 2: ...2...
Page 8: ...8...
Page 9: ...9 Sample Procedure Explained...
Page 10: ...10 Sample Procedure Explained continued...
Page 11: ...11 Sample Procedure Explained continued...
Page 12: ...12 Sample Procedure Explained continued...
Page 41: ...41 CHEMICAL ANALYSIS INFORMATION continued Figure 10 Multiple Standard Additions Graph...
Page 69: ...69 SECTION 4 PROCEDURES...
Page 70: ...70...
Page 92: ...92...
Page 100: ...100...
Page 108: ...108...
Page 116: ...116...
Page 139: ...139...
Page 140: ...140...
Page 168: ...168...
Page 176: ...176...
Page 184: ...184...
Page 194: ...194...
Page 202: ...202...
Page 214: ...214...
Page 234: ...234...
Page 242: ...242...
Page 264: ...264...
Page 270: ...270...
Page 284: ...284...
Page 298: ...298...
Page 314: ...314...
Page 324: ...324...
Page 332: ...332...
Page 336: ...336...
Page 340: ...340...
Page 354: ...354...
Page 370: ...370...
Page 378: ...378...
Page 386: ...386...
Page 398: ...398...
Page 404: ...404...
Page 412: ...412...
Page 424: ...424...
Page 436: ...436...
Page 446: ...446...
Page 456: ......
Page 459: ......
Page 460: ...460...