
49
CHEMICAL ANALYSIS INFORMATION
, continued
If you are using a Reagent Blank Correction, the blank correction should
be entered before the standard curve is adjusted.
To adjust the standard curve:
1.
Prepare the standard.
2.
Use the standard as the sample in the procedure.
3.
When the reading for the standard is obtained, press
SETUP
.
4.
Use the arrow keys to scroll to the “
STD
” setup option.
5.
Press
ENTER
to activate the standard adjust option.
6.
Edit the standard concentration to match that of the standard used.
7.
Press
ENTER
. A small plot of a line through a point will be displayed,
indicating that the curve has been adjusted with the standard.
Note:
If the attempted correction is outside the allowable adjustment limit, the
instrument will beep and flash Ø and the operation will not be allowed.
Preparing a User-Entered Calibration Curve
1.
Prepare five or more standards of known concentration that cover the
expected range of the test. Run tests as described in the procedure on
each prepared standard. Pour the customary volume of each known
solution into a separate clean sample cell of the type specified for
your instrument.
2.
Standardize (zero) the instrument using an untreated water sample or
a reagent blank, whichever the procedure instructs you to use.
3.
Measure and record the absorbance or %T of the known solutions. To
use %T vs. concentration see
%T Versus Concentration Calibration
.
To use absorbance vs. concentration, see
Absorbance Versus
Concentration Calibration
. Or create a user-entered program by
storing a custom calibration in the non-volatile memory of the
instrument. Refer to the section on entering user-entered programs in
the instrument manual.
%T Versus Concentration Calibration
If measuring %T, use semilogarithmic graph paper and plot %T (vertical
scale) versus concentration (horizontal scale). In
Figure 11
, iron standard
solutions of 0.1, 0.2, 0.4, 0.8, 1.2, 1.6, and 2.0 mg/L were measured on a
spectrophotometer at 500 nm using half-inch test tubes. Results were
plotted and the calibration table values were extrapolated from the
curve (
Table 7
).
Summary of Contents for DR/850
Page 2: ...2...
Page 8: ...8...
Page 9: ...9 Sample Procedure Explained...
Page 10: ...10 Sample Procedure Explained continued...
Page 11: ...11 Sample Procedure Explained continued...
Page 12: ...12 Sample Procedure Explained continued...
Page 41: ...41 CHEMICAL ANALYSIS INFORMATION continued Figure 10 Multiple Standard Additions Graph...
Page 69: ...69 SECTION 4 PROCEDURES...
Page 70: ...70...
Page 92: ...92...
Page 100: ...100...
Page 108: ...108...
Page 116: ...116...
Page 139: ...139...
Page 140: ...140...
Page 168: ...168...
Page 176: ...176...
Page 184: ...184...
Page 194: ...194...
Page 202: ...202...
Page 214: ...214...
Page 234: ...234...
Page 242: ...242...
Page 264: ...264...
Page 270: ...270...
Page 284: ...284...
Page 298: ...298...
Page 314: ...314...
Page 324: ...324...
Page 332: ...332...
Page 336: ...336...
Page 340: ...340...
Page 354: ...354...
Page 370: ...370...
Page 378: ...378...
Page 386: ...386...
Page 398: ...398...
Page 404: ...404...
Page 412: ...412...
Page 424: ...424...
Page 436: ...436...
Page 446: ...446...
Page 456: ......
Page 459: ......
Page 460: ...460...