390
Interferences
Summary of Method
In a highly acidic solution, ammonium molybdate reacts with
orthophosphate to form molybdophosphoric acid. This complex
is then reduced by the amino acid reagent to yield an intensely
colored molybdenum blue compound.
REQUIRED REAGENTS
Cat. No.
High Range Reactive Phosphorus Reagent Set (100 Test) ........................................... 22441-00
Interfering Substance
Interference Levels and Treatments
Calcium (Ca
2+
)
Greater than 10,000 mg/L as CaCO
3
Chloride
Greater than 150,000 mg/L as Cl
-
Colored samples
Add 1 mL of 10 N Sulfuric Acid Standard Solution to another 25-mL sample. Use
this instead of untreated sample as the blank to zero the instrument. Use a pipet
and pipet filler to measure the sulfuric acid standard.
High salt levels (Na
+
)
May cause low results. To eliminate this interference, dilute the sample until two
successive dilutions yield about the same result.
Magnesium
Greater than 40,000 mg/L as CaCO
3
Nitrites (NO
2
-
)
Bleach the blue color. Remove nitrite interference by adding 0.05 g of sulfamic
acid to the sample. Swirl to mix. Continue with Step 4.
Phosphates, high levels
(PO
4
3-
)
As the concentration of phosphate increases, the color changes from blue to green,
then to yellow and finally to brown. The brown color may suggest a concentration as
high as 100,000 mg/L PO
4
3-
. If a color other than blue is formed, dilute the sample and
retest.
Sulfide (S
2-
)
Sulfide interferes. For samples with sulfide concentration less than 5 mg/L,
sulfide interference may be removed by oxidation with Bromine Water as follows:
1.
Measure 50mL of sample into a 125-mL flask.
2.
Add Bromine Water dropwise with constant swirling until permanent yellow
color develops.
3.
Add Phenol Solution dropwise until the yellow color just disappears. Use this
sample in Steps 3 and 7.
Temperature
For best results, sample temperature should be
21 ±3 °C (70 ±5 °F).
Turbidity
May give inconsistent results for two reasons. Some suspended particles may
dissolve because of the acid used in the test. Also, desorption of orthophosphate
from particles may occur. For highly turbid samples, add 1 mL of 10 N Sulfuric
Acid Standard Solution to another 25-mL sample. Use this instead of untreated
sample as the blank to zero the instrument. Use a pipet and pipet filler to measure
the sulfuric acid standard.
Highly buffered samples or
extreme sample pH
May exceed the buffering capacity of the reagents and require sample
pretreatment.
PHOSPHORUS, REACTIVE,
continued
Summary of Contents for DR/850
Page 2: ...2...
Page 8: ...8...
Page 9: ...9 Sample Procedure Explained...
Page 10: ...10 Sample Procedure Explained continued...
Page 11: ...11 Sample Procedure Explained continued...
Page 12: ...12 Sample Procedure Explained continued...
Page 41: ...41 CHEMICAL ANALYSIS INFORMATION continued Figure 10 Multiple Standard Additions Graph...
Page 69: ...69 SECTION 4 PROCEDURES...
Page 70: ...70...
Page 92: ...92...
Page 100: ...100...
Page 108: ...108...
Page 116: ...116...
Page 139: ...139...
Page 140: ...140...
Page 168: ...168...
Page 176: ...176...
Page 184: ...184...
Page 194: ...194...
Page 202: ...202...
Page 214: ...214...
Page 234: ...234...
Page 242: ...242...
Page 264: ...264...
Page 270: ...270...
Page 284: ...284...
Page 298: ...298...
Page 314: ...314...
Page 324: ...324...
Page 332: ...332...
Page 336: ...336...
Page 340: ...340...
Page 354: ...354...
Page 370: ...370...
Page 378: ...378...
Page 386: ...386...
Page 398: ...398...
Page 404: ...404...
Page 412: ...412...
Page 424: ...424...
Page 436: ...436...
Page 446: ...446...
Page 456: ......
Page 459: ......
Page 460: ...460...