167
operator obtained a standard deviation of ±1.2 mg/L
cyanuric acid.
Estimated Detection Limit
The estimated detection limit for program 24 is 7.0 mg/L
cyanuric acid. For more information on the estimated detection
limit, see
Section 1
.
Interferences
Turbidity will interfere. Filter turbid samples before running the
test.
Summary of Method
The test for cyanuric acid uses the turbidimetric method.
Cyanuric Acid 2 Reagent precipitates any cyanuric acid present
and holds it in suspension. The amount of turbidity caused by the
suspended particles is directly proportional to the amount of
cyanuric acid present. Due to the nature of the precipitation
reaction, low levels of cyanuric acid (less than 7 mg/L) are not
detected by this method.
REQUIRED REAGENTS AND APPARATUS
Quantity Required
Description
Per Test
Unit
Cat. No.
Cyanuric Acid 2 Reagent Powder Pillow................ 1 pillow..................50/pkg ............2460-66
Sample Cell, 10-20-25 mL, w/cap ................................2 .........................6/pkg ..........24019-06
OPTIONAL REAGENTS
Cyanuric Acid ............................................................................................. 25 g ........... 7129-24
Water, deionized ........................................................................................... 4 L ............. 272-56
OPTIONAL APPARATUS
Balance, Acculab UI Series.........................................................................each ..........26947-00
Filter Paper, folded 12.5 cm ..................................................................100/pkg ............1894-57
Flask, volumetric, Class A, 100 mL............................................................each ......... 14574-42
Flask, volumetric, Class A, 1000 mL..........................................................each ......... 14574-53
Funnel, poly, 65 mm....................................................................................each ........... 1083-67
Pipet, Bulb...................................................................................................each ..........14651-00
Pipet, volumetric, Class A, 2.00 mL ...........................................................each ..........14515-36
For Technical Assistance, Price and Ordering
In the U.S.A.—Call 800-227-4224
Outside the U.S.A.—Contact the Hach office or distributor serving you.
CYANURIC ACID,
continued
Summary of Contents for DR/850
Page 2: ...2...
Page 8: ...8...
Page 9: ...9 Sample Procedure Explained...
Page 10: ...10 Sample Procedure Explained continued...
Page 11: ...11 Sample Procedure Explained continued...
Page 12: ...12 Sample Procedure Explained continued...
Page 41: ...41 CHEMICAL ANALYSIS INFORMATION continued Figure 10 Multiple Standard Additions Graph...
Page 69: ...69 SECTION 4 PROCEDURES...
Page 70: ...70...
Page 92: ...92...
Page 100: ...100...
Page 108: ...108...
Page 116: ...116...
Page 139: ...139...
Page 140: ...140...
Page 168: ...168...
Page 176: ...176...
Page 184: ...184...
Page 194: ...194...
Page 202: ...202...
Page 214: ...214...
Page 234: ...234...
Page 242: ...242...
Page 264: ...264...
Page 270: ...270...
Page 284: ...284...
Page 298: ...298...
Page 314: ...314...
Page 324: ...324...
Page 332: ...332...
Page 336: ...336...
Page 340: ...340...
Page 354: ...354...
Page 370: ...370...
Page 378: ...378...
Page 386: ...386...
Page 398: ...398...
Page 404: ...404...
Page 412: ...412...
Page 424: ...424...
Page 436: ...436...
Page 446: ...446...
Page 456: ......
Page 459: ......
Page 460: ...460...