
395
Sampling and Storage
Collect samples in clean plastic bottles. Analyze samples as soon
as possible after collection. If prompt analysis is not possible,
store samples for up to 28 days by cooling to 4 °C (39 °F) or
below. Warm samples to room temperature before analysis.
Accuracy Check
Standard Additions Method
a)
Open a Silica Standard Solution Bottle, 25 mg/L SiO
2
.
b)
Using the TenSette Pipet, add 0.1, 0.2, and 0.3 mL of
standard to three 10-mL samples. Mix thoroughly.
c)
Analyze each sample as described above. The silica
concentration should increase 0.25 mg/L for each 0.1 mL
of standard added.
d)
If these increases do not occur, see
Standard Addition
s in
Section 1
for more information.
Standard Adjust
To adjust the calibration curve using the reading obtained with
the
1.00-mg/L Standard Solution (see
Optional Reagents
)
, press the
SETUP
key and scroll (using the arrow keys) to the STD setup
option. Press
ENTER
to activate the standard adjust option. Then
enter
1.00
to edit the standard concentration to match that of the
standard used. Press
ENTER
to complete the adjustment. See
Section 1
,
Standard Curve Adjustmen
t for more information.
Method Performance
Precision
In a single laboratory, using standard solutions of 1.00 mg/L
silica and two representative lots of reagent and a instrument, a
single operator obtained a standard deviation of ±0.025 mg/L
silica.
Estimated Detection Limit (EDL)
The estimated detection limit for program 90 is 0.020 mg/L SiO2.
For more information on the estimated detection limit, see
Section 1
. If testing for very low levels of silica, use the ultra-low
range silica method on the Hach DR/2010 or DR/4000
Spectrophotometers.
SILICA, Low Range,
continued
Summary of Contents for DR/850
Page 2: ...2...
Page 8: ...8...
Page 9: ...9 Sample Procedure Explained...
Page 10: ...10 Sample Procedure Explained continued...
Page 11: ...11 Sample Procedure Explained continued...
Page 12: ...12 Sample Procedure Explained continued...
Page 41: ...41 CHEMICAL ANALYSIS INFORMATION continued Figure 10 Multiple Standard Additions Graph...
Page 69: ...69 SECTION 4 PROCEDURES...
Page 70: ...70...
Page 92: ...92...
Page 100: ...100...
Page 108: ...108...
Page 116: ...116...
Page 139: ...139...
Page 140: ...140...
Page 168: ...168...
Page 176: ...176...
Page 184: ...184...
Page 194: ...194...
Page 202: ...202...
Page 214: ...214...
Page 234: ...234...
Page 242: ...242...
Page 264: ...264...
Page 270: ...270...
Page 284: ...284...
Page 298: ...298...
Page 314: ...314...
Page 324: ...324...
Page 332: ...332...
Page 336: ...336...
Page 340: ...340...
Page 354: ...354...
Page 370: ...370...
Page 378: ...378...
Page 386: ...386...
Page 398: ...398...
Page 404: ...404...
Page 412: ...412...
Page 424: ...424...
Page 436: ...436...
Page 446: ...446...
Page 456: ......
Page 459: ......
Page 460: ...460...