GE H
EALTHCARE
D
IRECTION
FC091194, R
EVISION
11
V
IVID
7 S
ERVICE
M
ANUAL
Chapter 8 - Replacement Procedures
8 - 53
8-8-12-3
Verify TCP/IP Settings - Software Version v3.x
1.) From the top of the connectivity tab, select the
Tcpip
tab.
2.) Verify that all settings on the TCPIP tab matches those recorded in
Table 8-11 "Record settings from
Tcpip screen (v3.x)" on page 8-44
.
3.) Select
Save settings
.
4.) Select
Ok
to continue.
8-8-12-4
DC Offset Calibration
1.) Disconnect all probes.
2.) From the upper section of the window, select
Test
.
3.) Then select
Go to System Test.
4.) From the Module selection block, select
DC Offs. Cal
.
5.) Reboot the System when the DC Offset Calibration has been completed.
8-8-12-5
Verify System Software Version
1.) Press
CONFIG
(
F2
) and log on as
adm
, see
4-2-4 "Log On to the System as ‘ADM’" on page 4-12
2.) Select
about
tab.
3.) Verify that the SW versions are the same as what were recorded in
and Hardware versions (v3.x)" on page 8-46
.
8-8-13
Verification - Functional Checks
•
Please perform the tests specified in
Section 8-20 "Verification - Functional Checks" on page 8-105
Figure 8-48 TCP/IP Example Settings
Summary of Contents for Vivid 7
Page 1: ...GE Healthcare Operating Documentation Vivid 7 Service Manual Part Number FC091194 Revision 11...
Page 2: ......
Page 9: ...GE HEALTHCARE DIRECTION FC091194 REVISION 11 VIVID 7 SERVICE MANUAL vii JA ZH CN KO...
Page 38: ...GE HEALTHCARE DIRECTION FC091194 REVISION 11 VIVID 7 SERVICE MANUAL xxxvi...
Page 856: ...GE HEALTHCARE DIRECTION FC091194 REVISION 11 VIVID 7 SERVICE MANUAL Index 8...
Page 857: ......