Section 2-2 - General Console Requirements
2-2-2-3
Inrush Current
Inrush current is not a factor to consider due to the inrush current limiting properties of the power
supplies.
Maximum inrush current at startup:
•
230 VAC: 12 A
•
120 VAC: 7 A
2-2-2-4
Site Circuit Breaker
It is recommended that the branch circuit breaker for the unit be readily accessible.
2-2-2-5
Site Power Outlets
A dedicated AC power outlet must be within reach of the unit without extension cords. Other outlets
adequate for the external peripherals, medical and test equipment needed to support this unit must also
be present within 1 m (3.2 ft.) of the unit. Electrical installation must meet all current local, state, and
national electrical codes.
2-2-2-6
Unit Power Plug
If the unit arrives without a power plug, or with the wrong plug, you must contact your GE dealer or the
installation engineer must supply what is locally required.
2-2-2-7
Power Stability Requirements
Voltage drop-out
Max 10 ms.
Power Transients
(All applications)
Less than 25% of nominal peak voltage for less than 1 millisecond for any type of transient, including
line frequency, synchronous, asynchronous, or aperiodic transients.
FC000210
Vivid 7 (BT ’02)
100-120 VAC
±10%
10 A
50-60 Hz
FC000200
Vivid 7 (BT ’02)
230 VAC
±10%
5 A
50-60 Hz
FC000190
Vivid 7 PRO (BT ’02)
100-120 VAC
±10%
10 A
50-60 Hz
FC000180
Vivid 7 PRO (BT ’02)
230 VAC
±10%
5 A
50-60 Hz
FC000060
Vivid 7 (BT ’01)
100-120 VAC
±10%
10 A
50-60 Hz
FC000030
Vivid 7 (BT ’01)
230 VAC
±10%
5 A
50-60 Hz
CAUTION
POWER OUTAGE MAY OCCUR. The Vivid 7 requires a dedicated single branch circuit. To avoid
circuit overload and possible loss of critical care equipment, make sure you DO NOT have any
other equipment operating on the same circuit.
Table 2-4
Electrical Specifications for Vivid 7 (cont’d) sheet 3 of 3
GE VINGMED
PART NUMBER
DESCRIPTION
VOLTAGE
TOLERANCES
CURRENT
FREQUENCY
Summary of Contents for Vivid 7
Page 1: ...GE Healthcare Operating Documentation Vivid 7 Service Manual Part Number FC091194 Revision 11...
Page 2: ......
Page 9: ...GE HEALTHCARE DIRECTION FC091194 REVISION 11 VIVID 7 SERVICE MANUAL vii JA ZH CN KO...
Page 38: ...GE HEALTHCARE DIRECTION FC091194 REVISION 11 VIVID 7 SERVICE MANUAL xxxvi...
Page 856: ...GE HEALTHCARE DIRECTION FC091194 REVISION 11 VIVID 7 SERVICE MANUAL Index 8...
Page 857: ......