GE H
EALTHCARE
D
IRECTION
FC091194, R
EVISION
11
V
IVID
7 S
ERVICE
M
ANUAL
5 - 50
Section 5-3 - a Front-End Processor (FEP)
RF & Tissue Processor Board, RFT (cont’d)
5-3-18-7
Fuses and Dip-switches
None
5-3-18-8
LEDs and Jumpers
Figure 5-33 RFT, LEDs and Jumpers
Table 5-40 LEDs on the RFT Board
LED #
LED COLOR
DESCRIPTION
NORMAL STATE
LD8
- Red
Board Failure
OFF
LD7
- Green
Board OK
ON
LD6
- Yellow
DSP Runs
BLINKING
LD5
- Yellow
Data Out on PipeLink
ON During Data Transfers to PipeLink
LD4
- Yellow
Spare
OFF
LD3
- Red
Clock Problems
OFF
LD2
- Red
Clock Problems
OFF
LD1
- Red
Clock Problems
OFF
LD1
LD2
LD3
LD4
LD5
LD6
LD7
LD8
P5
1
2
39
40
P4
1
2
11
12
SW800