Chapter 8 - Replacement Procedures
8 - 27
8-7-12-4
Loading Application Software v7.x on Vivid 7 with BEP4.3
NOTE:
This procedure describes how to reload software that already has been stored as compressed
files on the BEP4.3’s hard disk drive. A CD is not used in this procedure.
To reload application software, follow this procedure:
1.) Reboot the Vivid 7, and press the
ESC
-key each second. The boot sequence will be interrupted and
a boot-menu appears on the screen.
There are two entries in the menu.
1.) Normal startup
2.) Recovery
Use the arrow keys to select (highlight) your choice.
2.) Select
Recovery.
3.) Press
Enter
to display the
System Recovery Menu
on the screen, see next page.
Figure 8-23 Select Recovery
Normal Startup
Recovery
GNU GRUB version 8.95 (639k lower / 523112X upper memory
Use the ^ and v keys to select which entry is highlighted.
Press enter to boot the selected OS or “p” to enter a
Password to unlock the next set of features.