Chapter 3 - System Setup
3 - 63
3-9-14-4 Verify the Connection to a Device (cont’d)
NOTE:
If the Check button fails directly, the AE Title is probably wrong.
If the Check button fails after a long time (corresponding to the timeout), the IP address or Port
Number is probably wrong.
In the illustration below, the message indicates that the pinging of the server worked fine (line 1),
but the application on the server didn’t respond (line 2).
Figure 3-61 Warning - Check status
Figure 3-62 Verification result
A green check-mark
indicates that the test
passed.
A red cross, as
illustrated, indicates
that the connection
failed.