GE H
EALTHCARE
D
IRECTION
FC091194, R
EVISION
11
V
IVID
7 S
ERVICE
M
ANUAL
8 - 106
Section 8-20 - Verification - Functional Checks
"Lower Rear Cover
Replacements Procedure" on
page 8-10
4-2-2 "Power ON/ Boot UP" on page 4-3
4-2-3 "Power Shut Down" on page 4-7
3-6-5-1 "Connect a Probe" on page 3-21
4-3-5 "Performance Tests" on page 4-31
4-3-6 "2D Mode (B mode) Checks" on page 4-32
4-3-21-3 "Brakes and Direction Lock Checks" on page 4-80
"Software Loading Procedure -
Software Version v7.x, v6.x,
v5.x and v4.x" on page 8-12
4-4-1 "Software Configuration Checks" on page 4-82
Perform
"Basic Functional Checks" on page 8-105
3-7-1-4 "Language Selection" on page 3-30
3-7-1-5 "Online Manual Language Selection" on page 3-31
"Software Loading Procedure -
Software Version v3.x" on page
8-38
4-4-1 "Software Configuration Checks" on page 4-82
Perform
"Basic Functional Checks" on page 8-105
3-7-1-4 "Language Selection" on page 3-30
3-7-1-5 "Online Manual Language Selection" on page 3-31
"Software Loading Procedure -
Software Version v2.x" on page
8-54
4-4-1 "Software Configuration Checks" on page 4-82
Perform
"Basic Functional Checks" on page 8-105
3-7-1-4 "Language Selection" on page 3-30
3-7-1-5 "Online Manual Language Selection" on page 3-31
"AC Power Replacement
Procedure" on page 8-82
Perform
"Basic Functional Checks" on page 8-105
10-7-4 "Grounding Continuity" on page 10-14
10-7-5 "Chassis Leakage Current Test" on page 10-16
10-7-6 "Isolated Patient Lead (Source) Leakage–Lead to
Ground" on page 10-17
10-7-7 "Isolated Patient Lead (Source) Leakage–Lead to
Lead" on page 10-18
10-7-8 "Isolated Patient Lead (Sink) Leakage-Isolation Test"
on page 10-19
10-7-9 "Probe Leakage Current Test" on page 10-21
"DC Power Replacement
Procedure" on page 8-86
Perform
"Basic Functional Checks" on page 8-105
10-7-4 "Grounding Continuity" on page 10-14
10-7-5 "Chassis Leakage Current Test" on page 10-16
10-7-6 "Isolated Patient Lead (Source) Leakage–Lead to
Ground" on page 10-17
10-7-7 "Isolated Patient Lead (Source) Leakage–Lead to
Lead" on page 10-18
10-7-8 "Isolated Patient Lead (Sink) Leakage-Isolation Test"
on page 10-19
10-7-9 "Probe Leakage Current Test" on page 10-21
"TX Power Replacement
Procedure" on page 8-89
Perform
"Basic Functional Checks" on page 8-105
10-7-4 "Grounding Continuity" on page 10-14
10-7-5 "Chassis Leakage Current Test" on page 10-16
10-7-6 "Isolated Patient Lead (Source) Leakage–Lead to
Ground" on page 10-17
10-7-7 "Isolated Patient Lead (Source) Leakage–Lead to
Lead" on page 10-18
10-7-8 "Isolated Patient Lead (Sink) Leakage-Isolation Test"
on page 10-19
10-7-9 "Probe Leakage Current Test" on page 10-21
Table 8-22 Functional Test Matrix (cont’d) sheet 2 of 3
Section
FRU Replacement
Diagnostics
Leakage
Current
Functional Test
Summary of Contents for Vivid 7
Page 1: ...GE Healthcare Operating Documentation Vivid 7 Service Manual Part Number FC091194 Revision 11...
Page 2: ......
Page 9: ...GE HEALTHCARE DIRECTION FC091194 REVISION 11 VIVID 7 SERVICE MANUAL vii JA ZH CN KO...
Page 38: ...GE HEALTHCARE DIRECTION FC091194 REVISION 11 VIVID 7 SERVICE MANUAL xxxvi...
Page 856: ...GE HEALTHCARE DIRECTION FC091194 REVISION 11 VIVID 7 SERVICE MANUAL Index 8...
Page 857: ......