Section 6-7 - Front-End Alignment Procedure (Beamformer Calibration)
Front-End Alignment Procedure (cont’d)
NOTE:
The LEDs displayed in the below individual board sections are relevant when the system is in
the default setting (2D-mode).
8.) Restart Vivid 7.
NOTE:
Always reboot the scanner after Front-End Alignment is finished and verify proper operation.
NOTE:
Do Not select “Cal Reset” after performing calibration. This will destroy the file located on the
Back-End Processor, the image quality will not be optimized and a new calibration will be
needed.
Figure 6-3 DC Calibration of the Beam Former AD-converters Passed