GE H
EALTHCARE
D
IRECTION
FC091194, R
EVISION
11
V
IVID
7 S
ERVICE
M
ANUAL
Chapter 5 - Components and Functions (Theory)
5 - 53
Spectrum Doppler Processor Board, SDP (cont’d)
5-3-19-10
Outputs
5-3-19-11
DC Voltages
5-3-19-12
Fuses and Dip-switches
None
Table 5-43 Output Signals
INPUT
DESCRIPTION
CONNECTION TO:
PipeLink
Doppler Signals extracted
IMP Board
Audio Output
Analog Doppler (I and Q) signals to Operator Panel
speakers and VCR
Internal I/O
VME Bus
Data communication
IMP
Table 5-44 DC Input Voltages
INPUT
DESCRIPTION
CONNECTION FROM:
+ 5 Vd
DC output voltage, distributed via Motherboard
(Backplane).
The “d” indicates that this voltage is used for digital circuits.
DC Power Supply via Backplane
- 5 Va
DC output voltage, distributed via Motherboard
(Backplane).
The “a” indicates that this voltage is used for analog circuits.
DC Power Supply via Backplane
GND
Distributed via Motherboard (Backplane)
DC Power Supply via Backplane