3 - 102
Section 3-10 - Connectivity Setup - Software v3.x
3-10-10
Query/Retrieve (Q/R) Setup for Software v3.x
3-10-10-1
Overview
Query/Retrieve was introduced with the V3.0 software.
The Query/Retrieve function makes it possible to search for and retrieve DICOM data from a DICOM
server for further analysis on the Vivid 7.
NOTE:
You may have to set up Vivid 7 as a destination on the server.
3-10-10-2
Query/Retrieve Setup on the Vivid 7
1.) Press
CONFIG
(
F2
) and log on as
adm
, see
4-2-4 "Log On to the System as ‘ADM’" on page 4-12
2.) Select Connectivity
3.) Select Dataflow
4.) Select Query Retrieve from the Name pull down menu
Figure 3-107 Select Query Retrieve
NAME PULLDOWN
MENU
QUERY RETRIEVE