GE H
EALTHCARE
D
IRECTION
FC091194, R
EVISION
11
V
IVID
7 S
ERVICE
M
ANUAL
Chapter 5 - Components and Functions (Theory)
5 - 31
Receiver Board, RX-128 (cont’d)
5-3-14-5
DC Voltages
5-3-14-6
Input Signals
5-3-14-7
Output Signals
5-3-14-8
Fuses
None
5-3-14-9
Jumpers and Dip-switches
None
Table 5-20 DC Input Voltages, Plug P3 & P4
INPUT
DESCRIPTION
CONNECTED TO
+ 5 Va
DC output voltage, distributed via Motherboard
(Backplane).
The “a” indicates that this voltage is used for analog circuits
DC Power Supply via Backplane
- 5 Va
DC output voltage, distributed via Motherboard
(Backplane).
The “a” indicates that this voltage is used for analog circuits.
DC Power Supply via Backplane
+15 V
DC output voltage, distributed via Motherboard (Backplane)
DC Power Supply via Backplane
- 15 V
DC output voltages, distributed via Motherboard
(Backplane)
DC Power Supply via Backplane
GND
Distributed via Motherboard (Backplane)
DC Power Supply via Backplane
Table 5-21 Input Signals
INPUT
DESCRIPTION
CONNECTION FROM
XD1 - XD128
Signals containing the received back-scattered echoes
from the active probe.
Active Probe > Relay Board > XD Bus
Boards > RX128 board
PEDOF
Signals containing the received back-scattered echoes
from PEDOF probe
Pedof probe > Relay Board > XD Bus
Boards > RX128 board
ATGCVP/ATGCVN
TGC Control Voltages
FEC Board/RFI Board > FE Bus >
RX128 board
Test SIG
Analog test signal
FEC Board/RFI Board > FE Bus >
RX128 board
Table 5-22 Output Signals
OUTPUT
DESCRIPTION
CONNECTION TO
RX1 - RX128
128 RX signals (echoes)
Beam Former Boards via Backplane
Summary of Contents for Vivid 7
Page 1: ...GE Healthcare Operating Documentation Vivid 7 Service Manual Part Number FC091194 Revision 11...
Page 2: ......
Page 9: ...GE HEALTHCARE DIRECTION FC091194 REVISION 11 VIVID 7 SERVICE MANUAL vii JA ZH CN KO...
Page 38: ...GE HEALTHCARE DIRECTION FC091194 REVISION 11 VIVID 7 SERVICE MANUAL xxxvi...
Page 856: ...GE HEALTHCARE DIRECTION FC091194 REVISION 11 VIVID 7 SERVICE MANUAL Index 8...