Chapter 1 - - Safety and Service
Page 29
4.5.1
Personnel Requirements
4.5.2
Overview
This procedure can be performed with:
• Fluke tester
4.5.3
Tools and Test Equipment
Note:
The equivalent means that the tool is designed to be capable of performing tests to both the
standard of IEC 62353 and the earth leakage current of IEC 60601-1.
RISK OF ELECTRIC SHOCK.
VOLTAGE PRESENT.
THE SYSTEM MUST BE DISCONNECTED FROM THE FACILITY MAINS INCLUDING THE A1
PANEL GROUND WIRE.
4.5.4
Required Conditions
4.5.5
Set up tools
• How to connect the Tool
1). Connect power cord
2.) Connect red test lead to red input jack (V/Ω/A)
3.) Connect black test lead to black input jack
Personnel Requirements
Preliminary Reqs
Procedure
Finalization
1
Not Applicable
30min
10min
Item
Qty.
Effectivity
Part#
Manufacturer
Fluke ESA612 or equivalent
1
-
-
Fluke
Condition
Reference
Effectivity
All sub-system covers must be removed to access the
grounding points within each sub-system.
1
-