Chapter 5 - - Installation & Configuration
Page 57
5.2
Assembly Placement
Figure 5-9 AP Placement Locations - Control Room Outside
To place the AP in a proper location:
1.) Place the AP at a location 2.5 meters above the floor.
2.) No obstacles are between AP and application range of image detector.
3.) If the installation criteria for "Default Location 1" is not available, try “Default Location 3", then
“Default Location 2".
Note:
For all locations, the d
istance between AP and detector
should be
less than
4
meters
.
4.) If all three recommended installation points are all not available, follow the instructions in
Assembly Installation on page 122
to test the wireless signal quality in the adjusting range to
find a proper location to place the AP.
Table
WS
Default Location
Adjusting Range
Detector Insert Direction
Exam Room
< 4 met
ers
< 4 meters
< 4 met
ers
Default Location 1-1
(Recommended)
Default L
ocation 2
Default Location 3
Default Location 1-2
(Recommended)
* Default Location 1-1: Only table bucky is required.
* Default Location 1-2: If wallstand is required.