Section 11.0 - AP Assembly Installation
Section 11.0
AP Assembly Installation
11.1
AP Assembly location selection
11.1.1
AP location selection process
1.) Place the AP at Default location and then perform wireless signal strength test. If passed,
install AP at default location. Otherwise, continue Step 2.
Figure 5-21 Room Layout
2.) Place AP in Adjusting range and then perform wireless signal quality test. If passed, install AP
at selected location. If failed to find a good location within Adjusting range, select any possible
location where wireless signal is good. Place AP at the location, then perform wireless quality
test. Compare wireless quality test results of all passed locations, select the best one to install
AP.
Table
WS
Default Location
Adjusting Range
Detector Insert Direction
Exam Room
< 4 met
ers
< 4 meters
< 4 met
ers
Default Location 1-1
(Recommended)
Default L
ocation 2
Default Location 3
Default Location 1-2
(Recommended)
* Default Location 1-1: Only table bucky is required.
* Default Location 1-2: If wallstand is required.