Chapter 12 - - Set Preferences
Page 271
GE H
EALTHCARE
D
IRECTION
5461425-8EN, R
EVISION
B
RIVO
XR118 S
ERVICE
M
ANUAL
Click the detector Management icon [
] to list all the detectors registered
in the system.
To associate a “Non-associated”:
1. Selects a non-associated detector from the detector list which has been boot
up.
2. Clicks “Associate” button to make it as associated detector.
3. After the detector association is done. The detector is marked with asterisk
as associated detector. The prior associated detector changes to
nonassociated.
After detector associated with the system, if the detector FW version not
compatible, the system shall display FW incompatible message in the
system inhibit message bar.
When system uploads the detector calibration data, there shall be a
message indiates that detector calibration data upload is in progress in
the system inhibit message bar.
To identify a detector
1. Click the [Identify] button to confirm the wireless connection of the
associated detector.
[Identify] button will be active only when associated detector is
selected.
Prior to identifying the associated detector, the detector should be
successfully boot up and wirelessly connected to the system.
2. The associated Detector connection LED blinks for 15 seconds if
identification is successful.
If the associated detector is in sleep mode, the identify process will
indicate user that the associated detector is not available. Please refer
to
to wake up the detector.
If there is a detector communication error, the identify process will
indicate user of the identification failure.
To release an associated detector
1. Selects the associated detector from the detector list which has been boot
up.
When the associated detector is selected, the “Associate” button
changes to “Release”.
2. Click the [Release] button.
3. A pop up box will be displayed requiring your confirmation. Click [Yes] to
proceed.