GE H
EALTHCARE
D
IRECTION
5461425-8EN, R
EVISION
B
RIVO
XR118 S
ERVICE
M
ANUAL
Chapter 5 - - Installation & Configuration
Page 121
To release an associated detector
1.) Selects the associated detector from the detector list which has been boot up.
•
When the associated detector is selected, the “Associate” button changes to “Release”.
2.) Click the [Release] button.
•
A pop up box will be displayed requiring your confirmation. Click [OK] to proceed.
3.) Upon releasing the associated detector successfully, associated detector will be updated as
non-associated.
•
If the detector release failed, the associated detector will not become non-associated. A popup
message will be displayed to indicate user of the release failure.
10.4
Detector Sleep Time
Figure 5-19 Detector Sleep Time
The Digital Detector can be set to go into the sleep mode at a designated time to save the battery
power of the Digital Detector. If the Digital Detector sleep time is set to 30 minutes, the Digital
Detector it will automatically go into the sleep mode after 30 minutes of NO activity on the system.
This includes any activity on the screen or movement of the system. Once the Digital Detector is in
sleep mode you must press the power button on the Digital Detector for one to two seconds to exit
the sleep mode. The Digital Detector will wake up and be ready for an exposure in 10 seconds.
10.5
Grid Line Rate
Figure 5-20 Grid Line Rate
Configure the grid line parameters on the target analog and enable Grid Line Reduction function if
required.