GE
V
OLUSON
i / V
OLUSON
e
D
IRECTION
KTI106052, R
EVISION
10
S
ERVICE
M
ANUAL
3-20
Section 3-4 - Preparing for Set Up
3-4-2
System Voltage Settings
NOTE:
Not necessary at the Voluson i / Voluson e portable ultrasound system:
For details see:
Table 2-3 "Electrical Specifications for Voluson i / Voluson e" on page 2-2
.
3-4-3
Modo Cart Voltage Configuration
NOTE:
Only required when using the Voluson i / Voluson e with the optional Modo Cart.
Check Fuses and Voltage Output Selector setting located on the isolation transformer of the Modo Cart.
If required switch Output Selector to correct position and exchange fuses of the isolation transformer.
a.) Open the fuse protection on the rear of the isolation transformer.
b.) Pull out the fuse holder and exchange the 2 fuses.
c.) Push the fuse holder back into position (correct voltage should be shown in upper line,
e.g, 230V, see:
Figure 3-16
, above) and then close protection.
CAUTION
!! CAUTION:
Before starting the system or peripherals connected to the Modo Cart, always make sure:
• Rating of isolation transformer fuses is compatible. •
Output Selector is in correct position.
The wrong fuses and position of the voltage output selector may cause major damage on
Voluson i / Voluson e and connected peripherals.
CAUTION
!! CAUTION:
Rating of fuses depend on Voltage (refer to description on the isolation transformer box).
Figure 3-16 Change fuses of isolation transformer
Isolation Transformer
Voltage
Fuses
Output
Selector
Fuse holder
(e.g. 230V)
Fuse
Fuse
Fuse holder
(back)
Summary of Contents for H48651KR
Page 2: ......
Page 11: ...GE VOLUSON i VOLUSON e DIRECTION KTI106052 REVISION 10 SERVICE MANUAL ix ZH CN KO ...
Page 44: ...GE VOLUSON i VOLUSON e DIRECTION KTI106052 REVISION 10 SERVICE MANUAL xlii Table of Contents ...
Page 514: ...GE VOLUSON i VOLUSON e DIRECTION KTI106052 REVISION 10 SERVICE MANUAL IV Index ...
Page 515: ......