Section 8-20 - Replacement of the Voluson Station Components
8-20-9
Replacement of the optional Probe Mux Box
8-20-9-1
Manpower
One person, 5 minutes
8-20-9-2
Tools
Phillips screwdriver 1
8-20-9-3
Preparation
1.) Power Off/Shutdown the system as described in
Section 3-6-3 on page 3-72
.
2.) Remove the Voluson i / Voluson e from the Tray.
8-20-9-4
Probe Mux Box - Removal Procedure
1.) Disconnect the Probe connector.
2.) Hold the Probe box and screw off the 4 screws.
Figure 8-88 disconnect the Probe connector
Figure 8-89 unscrew the 4 screws