Chapter 10 - Care & Maintenance
10-11
10-5-5-4
Disinfection by means of the RIC-Holder
Especially for Real-time 4D endocavity probes (RIC), it is necessary to take extreme care when
transporting the system with the probe attached, or during the disinfection process.
Inadequate handling may lead to dead elements, shocked head mechanics, etc.
The RIC-Holder (especially developed for RIC Real-time 4D endocavity probes) guarantees that the
sensitive probe head does not hit the bottom of the container during the disinfection procedure.
NOTE:
Instructions are supplied with each RIC-Holder (KTZ225469).
Section 10-6
Using a Phantom
The use of a Phantom is not required during Preventive Maintenance.
Customer may use it as part of their Quality Assurance Program tests.
OK
OK
OK
OK
RIC-Holder
Endocavity soaking cup kit
RIC-Holder
RIC-Holder