GE
V
OLUSON
i / V
OLUSON
e
D
IRECTION
KTI106052, R
EVISION
10
S
ERVICE
M
ANUAL
7-18
Section 7-7 - Minimum Configuration to Boot/Scan
7-7-1
Booting without using the Ultrasound Hardware
(cont’d)
7.) Save a “Full Backup” as described in
Section 4-5-3 on page 4-32
, or press the
ALT
+
D
key to
gather detailed information and data from the system, see:
Section 7-4-2-1 on page 7-9
.
8.) Press
UTILITIES
, choose SYSTEM, select the
SYSTEM INFO
page and click REBOOT MACHINE.
Figure 7-22 Reboot Machine