Section 8-20 - Replacement of the Voluson Station Components
8-20-19
Replacement of the Top Shelf Bottom Front Cover
8-20-19-1
Manpower
One person, 10 minutes
8-20-19-2
Tools
Phillips screwdriver 1
8-20-19-3
Preparations
1.) Power Off/Shutdown the system as described in
Section 3-6-3 on page 3-72
.
2.) Remove the Gel Condom Holder as described in
Section 8-20-4 on page 8-63
or
remove he optional Probe Holder for Endocavity Probes
as described in
Section 8-20-5 on page 8-64
.
8-20-19-4
Top Shelf Bottom Front Cover - Removal Procedure
1.) Screw off the 5 screws around the Top Shelf Bottom Front Cover.
2.) Screw off the 2 screws by the USB port.
8-20-19-5
Top Shelf Bottom Front Cover - Installation Procedure
1.) Place the new Top Shelf Bottom Front Cover, fasten the 2 screws of the USB-port and fasten the 5
screws around it.
2.) Mount the Gel Condom Holder as described in
Section 8-20-4-5 on page 8-63
or
mount the optional Probe Holder for Endocavity Probes
as described in
Section 8-20-5-5 on page 8-64
.
Figure 8-101 unscrew the screws