Chapter 7 - Diagnostics/Troubleshooting
7-31
File Type Unknown
load volume files from other storage medium
File Volume size not consistent
load volume files from other storage medium
FilterBuffer memory allocation failed
restart the system
FLT: Denormal operand
restart the system
FLT: Divide by zero
restart the system
FLT: Invalid operation
restart the system
FLT: Overflow
restart the system
FLT: Stack overflow
restart the system
FLT: Underflow
restart the system
GeoDescription3D_TissueCF::operator=: handed over
argument is not of same type, dynamic cast failed!!
restart the system
GeoDescriptionMotion Constructor: unknown mode Parameter
handed over
restart the system
GeoDescriptionMotion::operator=: handed over argument is not
of same type, dynamic cast failed!!
restart the system
GeoDescriptionPWMode::operator=: handed over argument is
not of same type, dynamic cast failed!!
restart the system
GeoDescriptionTissueCF Constructor: unknown mode
Parameter handed over
restart the system
GeoDescriptionTissueCF::operator=: handed over argument is
not of same type, dynamic cast failed!!
restart the system
Hardware doesn't support CW-mode
pencil probe + CW-Hardware not available->HW problem
HardwareRelatedSoftware_Windows in write have different
ProbeAcousticUnitIDs
disconnect all connected probes and connect them again;
if not ok restart the system
HardwareRelatedSoftware_Windows in write have different
ProbeScanFuncIDs
disconnect all connected probes and connect them again;
if not ok restart the system
IBegrenzer.cpp BCshots no TxPulseEntrys
restart the system
IBegrenzer.cpp Bshots no TxPulseEntrys
restart the system
IBegrenzer.cpp Bshots TxMultiFocus problem
restart the system
IBegrenzer.cpp CWshots no TxPulseEntrys
restart the system
IBegrenzer.cpp MCshots no TxPulseEntrys
restart the system
IBegrenzer.cpp Mshots no TxPulseEntrys
restart the system
IBegrenzer.cpp Mshots TxMultiFocus problem
restart the system
IBegrenzer.cpp PWshots no TxPulseEntrys
restart the system
In the 3D Image Measure is not allowed
change to another format than 3D Fullscreen mode
Incorrect ReplayCtrl-Object
restart the system
Error Messages
Actions
Summary of Contents for H48651KR
Page 2: ......
Page 11: ...GE VOLUSON i VOLUSON e DIRECTION KTI106052 REVISION 10 SERVICE MANUAL ix ZH CN KO ...
Page 44: ...GE VOLUSON i VOLUSON e DIRECTION KTI106052 REVISION 10 SERVICE MANUAL xlii Table of Contents ...
Page 514: ...GE VOLUSON i VOLUSON e DIRECTION KTI106052 REVISION 10 SERVICE MANUAL IV Index ...
Page 515: ......