Section 10-7 - Electrical Safety Tests
10-7-7
Chassis Leakage Current Test
10-7-7-1
Definition
This test measures the current that would flow through a grounded person who touches the accessible
conductive parts of the equipment during normal and fault conditions.
The test verifies the isolation of the power line from the chassis.
The meter is connected to parts of the equipment, easily contacted by the user or patient.
Measurements should be made under the test conditions specified in:
•
Table 10-10 on page 10-13
, or
•
Table 10-11 on page 10-13
, as applicable.
Record the highest reading.
10-7-7-2
Generic Procedure
The test verifies the isolation of the power line from the chassis.
The testing meter is connected from accessible metal parts of the case to ground.
Measurements should be made under the test conditions specified in:
•
Table 10-10 on page 10-13
, or:
•
Table 10-11 on page 10-13
, as applicable.
Record the highest reading of current.
When using the Microguard or a similar test instrument, its power plug may be inserted into the wall
outlet and the equipment under test is plugged into the receptacle on the panel of the meter.
This places the meter in the grounding conductor and the current flowing from the case to ground will
be indicated in any of the current ranges.
The maximum allowable limit for chassis source leakage is shown in:
•
Table 10-10 on page 10-13
, or
•
Table 10-11 on page 10-13
, as Chassis/Enclosure Leakage.
DANGER
!! DANGER:
Electric Shock Hazard.
When the meter's ground switch is OPEN, DO NOT touch the unit!
CAUTION
!! CAUTION:
Equipment damage possibility. Never switch the Polarity and the status of Neutral when the unit
is powered ON. Be sure to turn the unit power OFF before switching them using the POLARITY
switch and/or the NEUTRAL switch. Otherwise, the unit may be damaged.
Summary of Contents for H48651KR
Page 2: ......
Page 11: ...GE VOLUSON i VOLUSON e DIRECTION KTI106052 REVISION 10 SERVICE MANUAL ix ZH CN KO ...
Page 44: ...GE VOLUSON i VOLUSON e DIRECTION KTI106052 REVISION 10 SERVICE MANUAL xlii Table of Contents ...
Page 514: ...GE VOLUSON i VOLUSON e DIRECTION KTI106052 REVISION 10 SERVICE MANUAL IV Index ...
Page 515: ......