Chapter 8 - Replacement Procedures
8-13
8-2-2
FMI from external device (available at Software 8.x.x)
8-2-2-1
Introduction
8-2-2-2
Manpower
One person ~ 1 hour (depends on contents of USB System Stick, peripherals, etc.)
8-2-2-3
Tools
USB System Stick
8-2-2-4
Preparations
Before performing the Software Update/Upgrade:
A.) make sure that all system functions are working correct
B.) check the current Ultrasound Application Software version and the installed Options
NOTE:
It is
NOT necessary
to save Full System Configuration (Full Backup) prior to the upgrade.
All existing User Programs, 3D/4D Programs and Auto Text settings remain untouched!
NOTICE
!! NOTICE:
To update the Systems C:\ image via the FMI FROM EXTERNAL DEVICE button in the System Setup
SERVICE
page,see:
Section 8-2-2-5 "System Software - Installation Procedure (FMI from external
device)" on page 8-16
.
The Software parts to be upgraded (e.g., Ultrasound Application Software, Service Software, EUM,
MS Patches, etc.) depend on the contents of the USB System Stick being used.
NOTICE
f the currently installed software has to be upgraded to a newer version, and the system DVD being
used is
SW 8.2.0 or higher
, a new software specific “Permanent key” is required.
(Calculated in OKOS
Option Key Order System;
http://3.187.187.9/OKOS.
If you do not have permission to access OKOS, contact the
O
n
L
ine
C
enter to get the necessary key.
Summary of Contents for H48651KR
Page 2: ......
Page 11: ...GE VOLUSON i VOLUSON e DIRECTION KTI106052 REVISION 10 SERVICE MANUAL ix ZH CN KO ...
Page 44: ...GE VOLUSON i VOLUSON e DIRECTION KTI106052 REVISION 10 SERVICE MANUAL xlii Table of Contents ...
Page 514: ...GE VOLUSON i VOLUSON e DIRECTION KTI106052 REVISION 10 SERVICE MANUAL IV Index ...
Page 515: ......