Functional check
59
Document no. M1144951-004
5.3.6 Compact Airway Module, E-CXXXXX/ M-CXXXXX
Wait until the message ‘Calibrating gas sensor’ disappears from the screen.
1.
Check that the fan is running.
2.
If the module contains membrane keys on the front panel, press each of the keys for at
least one second and check that they are identified.
"
Compact Airway Modules with the Patient Spirometry option
3.
Connect a clean spirometry tube and D-lite to the module. Connect the sampling line.
Breathe through the wider side of the D-lite.
Check that the flow waveform moves downwards when you breathe in and upwards
when you breathe out.
"
For all Compact Airway Modules
4.
Block the tip of the sampling line with your finger and check that the message ‘Sample
line blocked’ appears on the monitor screen within 30 seconds.
5.
Detach the D-fend and check that the message ‘Check D-fend’ appears on the monitor
screen within 30 seconds.
6.
Breathe to the sampling line briefly. Check that the CO2 information is updated on the
screen.
"
5.3.7 Single width Airway Module, E-miniC/ M-miniC
Wait until the message ‘Calibrating gas sensor’ disappears from the screen.
1.
Block the tip of the sampling line with your finger and check that the message ‘Sample
line blocked’ appears on the monitor screen within 30 seconds.
2.
Detach the Mini D-fend and check that the message ‘Check D-fend’ appears on the
monitor screen within 30 seconds.
Breathe to the sampling line briefly. Check that the CO2 information is updated on the
screen.
"
5.3.8 Tonometry Module, E-TONO/ M-TONO
1.
Press each of the front panel membrane keys for at least one second and check that they
are identified.
"
5.3.9 Multiparameter Hemodynamic Modules
ECG and RESP measurements
1.
Check the function of the front panel membrane key
ECG
,
Lead
, if available.