Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
18
Document no
.
M1144953-001
21. Attach an adult NIBP hose and cuff onto your arm and perform one NIBP measurement.
Check that the module identifies the cuff, i.e. the text ‘Adult’ appears in the NIBP digit field
for a short time.
Check that the module gives a reasonable measurement result.
"
E-PSM(P), E-PRESTN:
22. Attach a NIBP cuff hose without cuff identification and check that the module identifies
the hose:
−
The message ‘Select inflation limits’ appears in the NIBP digit field.
−
When you try to start the measurement, the monitor automatically opens the
selections
NIBP Setup - Inflation Limits
.
"
2.2.10 EEG Module, E-EEG/ M-EEG and EEG Headbox, N-EEG
1.
Enter the
EEG & EP
service menu:
Check that the Time-outs, Bad checksums and Bad c-s by mod values in the module view
are not increasing faster than by 5 per second. Check that the memories of the module
have passed the internal memory test, i.e. RAM, ROM and EEPROM all state OK.
Check that the HB Mod Time-outs, HB Mod Bad Checksum and Mod HB Bad Checksum
values are not increasing faster than by 5 per second. Check that the memory of the
headbox has passed the internal memory test, i.e. HB Rom Error in the headbox view
states 0.
2.
Check the
EP Start/Stop
and
Imp. Check
membrane keys both in the module and the
headbox.
3.
Connect the EEG simulator to the headbox. Select 10 k
Ω
as imped. pos. and imped. neg.
value on the simulator. Go to
EEG & EP
service menu and select
Check Electr.
Check in the
headbox view that the impedances in all four channels are 10 k
Ω
±
1 k
Ω
.
4.
Select 2 k
Ω
as imped. pos. and imped. neg. value on the simulator. Select 10 Hz 200
μ
V
sinewave on the simulator and check that all the four waveforms have the same form.
Check that the size of the waveforms is 200
μ
V
pp
±
5 V. Check that the MF value is 10
±
0.5
Hz. Check that the amp value is 71
μ
V
±
3
μ
V.
5.
Plug in the earphones to the headbox.
NOTE: Be careful with loud stimulation from the earphones when starting AEP stimulation.
Start AEP stimulation by pressing the
EP Start/Stop
button on the module. Check that
the clicking sound comes from the earphones in 1.1 Hz frequency. Stop the stimulation by
pressing again the
EP Start/Stop
button on the module. Check that the clicking stopped.
"
2.2.11 BIS Module, E-BIS/ M-BIS
1.
Enter the
BIS
service menu:
Check that the DSC is connected to the module and the PIC+ cable is connected to the
DSC.