Display Unit for F-CM(C)1 (Rev. 05), F-CM(C)REC1 (Rev. 01)
29
Document no. M1144956-003
39. Data Card recognition
Insert a Memory card labelled Data in the right hand side memory card slot. Check that
the message ‘Data Card inserted’ appears in the message field and the green menu card
symbol on the upper right hand corner of the screen within 1 minute.
Wait until the information regarding SLOT2 is fully updated in the service menu, then
check that the Card type states DATA and the File system states ATA.
Check that the rest of the information for SLOT2 is reliable and no errors have been
detected.
"
40. Save and Load Modes
Enter the
Save Modes
menu:
Monitor Setup
Install Service
(Password 16-4-34)
Save Modes
(Password 13-20-31)
Save the current modes into the Menu card by selecting
Load Modes
To Memory Card --> Save.
Wait until the text ‘Saved’ appears, then return to the previous menu.
Change the name for the mode number 1:
Select the mode number 1, press the ComWheel and select
-Name.
Select suitable
characters from the list by turning and pressing the ComWheel, then confirm the new
name by selecting
End.
Load the original modes from the Menu card by selecting
Load Modes
From Memory Card --> Load
.
Wait until the text ‘Loading’ changes to ‘Loaded’ (may take a couple of minutes), then
return to the previous menu.
Check that the mode number 1 has got back its original name.
"
41. Data Card functions
Press the membrane key
Display Trends
or
Pt.Data & Trends
--> Trends
. Check that
there is trend information available for the monitored parameters.
Erase the trends:
Reset Case
Reset All --> Yes
or
Admit/Discharge
Discharge --> Yes
Check that the trends have been erased.
Reload the trends from the Data card by pressing the key
Patient Data
or
Admit/
Discharge
, selecting
Patient from Card
, pressing the ComWheel on the last saved file
(the file information is shown at the bottom of the menu) and selecting
Load
.