Service Menu
5
Document no. M1144955-001
Subnet Status
The Subnet status view gives more accurate
information of the different subnet id:s connected.
All four
Subnet status
menus have a similar
structure. The number of different packets
transmitted and received by the monitor are
shown in the columns below Tx and Rx. The packet
types are:
Waveforms:
Waveform data
Phys. data:
Physiological numerical data
Alarms:
Alarms, alarm profiles and alarm
limits
Link mgmt:
Network management messages
Record K:
Record Keeper data
MonToMon:
Monitor-to-monitor communication
related data
Printer:
Printing data and control messages
File Op.:
File operation messages, saving
and
loading of cases
Service:
Maintenance and service
Modes:
User mode data
Indics.:
Remote indications sent to monitor
RemoteEv:
Remote events
Data server:
Packets of the data server
(Arrhythmia Workstation)
Packets total:
Total number of packets
sent/received
Bytes total:
Total number of bytes
sent/received
T-o
Number or resendings
InE
Received faulty packets
LenE
Erroneous packet length
Dupl
Same packet received as a duplicate