System installation
27
Document no. M1144951-004
2.
Detach the cover plate on the network connectors at the rear of the frame unit (above the
power supply unit). Leave the collar on the connectors attached.
3.
Open the cover for card drive slots (on the left side of the monitor). Insert the option
software card into one of the card drive slots and press the software card firmly in
position.
4.
Switch the monitor on.
5.
Wait until the normal screen appears and check that a message 'Field upgrade OK' is
displayed as a note of successful upgrade.
6.
Remove the option software card.
7.
Enter the service menu and make sure that both options were loaded (
Monitor Setup
-Install/Service - Service - Frame - Options - "Net+DC"
).
8.
Install the device plate for option software on inside the cover for card drive slots. Close
the cover.
9.
Test the monitor with a MemCard and by connecting it to Datex-Ohmeda Network.
10. Fill out all necessary documentation regarding the upgrade.
3.3.10 Installing the Datex-Ohmeda Network Upgrade, U-CMNET
The following instructions apply to upgrading the monitor with the Network option. Detailed
upgrade instructions are supplied with the corresponding upgrade product.
NOTE: You can download the option software only onto one monitor. During the downloading
of option software, the serial number of the monitor is written on the software card, and if the
downloading for some reason would fail, the software can be downloaded again onto the
same monitor, but not onto any other monitor.
1.
Make sure that the monitor is switched to standby.
2.
Detach the cover plate on the network connectors at the rear of the frame unit (above the
power supply unit). Leave the collar on the connectors attached.
3.
Open the cover for card drive slots (on the left side of the monitor). Insert the option
software card into one of the card drive slots and press the software card firmly in
position.
4.
Switch the monitor on.
5.
Wait until the normal screen appears and check that a message 'Field Upgrade OK' is
displayed as a note of successful upgrade.
6.
Remove the option software card.
7.
Enter the service menu and make sure that the option was loaded (
Monitor Setup
-Install/Service - Service - Frame - Options - "Net"
).
8.
Install the device plate for option software on inside the cover for card drive slots. Close
the cover.
9.
Test the monitor by connecting it to Datex-Ohmeda Network.
10. Fill out all necessary documentation regarding the upgrade.
3.3.11 Installing the S/5 Wireless LAN Upgrade, U-CMW
NOTE: WLAN CF card with PCMCIA adapter is compatible only with Compact Monitor frame
F-CM(C)1..05 or later.
The following instructions apply to upgrading the monitor with the Wireless LAN option.
Detailed upgrade instructions are supplied with the corresponding upgrade product.
You can download the option software only onto one monitor. During the downloading of