Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
14
Document no
.
M1144958-001
•
the fuses are of the correct rating
Reconnect the fuse holder and check that it locks up properly.
"
3.2.5 Functional inspection
10. Stand-by LEDs
Connect the power cord.
Check that the standby LED starts blinking.
Check that the fan starts running.
"
11. Start-up
Switch the monitor on.
Check that the monitor starts up properly, i.e. the alarm LEDs turn blank, the start-up
sound is heard from the loudspeaker and the normal monitoring screen appears.
No error messages should appear on the screen.
"
12. Module recognition
Configure the monitor screen so that all the needed parameter information is shown:
Monitor Setup
Screen 1 Setup
Waveform Fields
Digit Fields
Check that the selected parameter information is shown on the screen.
NOTE: InvBP waveforms are not shown until without a patient simulator. Do no connect the
simulator to the monitor at this stage.
"
13. Main software
Enter the service menu:
Monitor Setup
Install/Service
(Password 16-4-34)
Service
(Password 26-23-8)
Take down the information regarding Power Management Board software and Battery.
"
14. Content of service log
Select
Service Log
from the menu. Check the contents of the Service Log for possible
problems. If the monitor contains the recorder unit, record the Service Log on the
recorder by selecting
Record Log
.
"