System installation
29
Document no. M1144951-004
7.
Attach the recorder to the side panel with two screws at the back of the recorder (the
screws can be accessed inside the recorder).
8.
Check that the recorder and the side panel are aligned and the recorder tightly attached.
Figure 8
Installation of the U-CMREC1
9.
Attach the device plate to the back cover of the monitor, near the other device plates.
3.4 S/5 Remote Controller, K-CREMCO
To connect Remote Controller, K-CREMCO, to a Compact Monitor:
1.
Make sure that the monitor is switched to standby.
2.
Connect the K-CREMCO to the 5-pin DIN connector X9 that is located on the right side of
the Compact Monitor display unit.
3.5 Anesthesia Record Keeping Keyboard, K-ARKB
To connect Anesthesia Record Keeping Keyboard, K-ARKB, to a Compact Monitor:
1.
Make sure that the monitor is switched to standby.
2.
Connect the K-ARKB with an appropriate cable to the 5-pin DIN connector X9 that is
located on the right side of the Compact Monitor display unit.
3.6 ARK Barcode Reader, N-SCAN (optional)
3.6.1 Connection to Compact Monitor
1.
Make sure that the monitor is switched to standby.
2.
Connect the N-SCAN with an appropriate cable to the 5-pin DIN connector X9 that is
located on the right side of the Compact Monitor display unit.