Frame Unit for F-CM(C)1 (Rev. 05), F-CM(C)REC1 (Rev. 01)
15
Document no. M1144958-001
15. Voltages
Check the power supply unit output voltages through the service menu:
Monitor Setup
Install/Service
(Password 16-4-34)
Service
(password 26-23-8)
Frame
Power Supply
The output voltages should meet the following ranges:
Preset monitor settings:
Record/Print
Record Waveforms
Waveform 1 --> ecg1
Waveform 2 --> p1
Waveform 3 --> p2
Delay --> Off
Paper Speed --> 6.25 mm/s
Length --> 30 s
16. Module communication
Connect a patient simulator to the module.
Enter the
ESTP : ECG
service menu:
Monitor Setup
Install/Service
(Password 16-4-34)
Service
(password 26-23-8)
Parameters
ESTP : ECG
Check that the Timeouts, Bad checksums and Bad c-s by mod values do not increase
faster than by 5 per second.
Change the module into the other slot. When the module is recognized, check that the
values do not still increase faster than by 5 per second.
"
17. Battery voltage
Enter the
Battery
service menu:
Monitor Setup
Install/Service
(password 16-4-34)
Service
(password 26-23-8)
+15 V
14.50...15.50 V
-15 V
-14.50...-15.50 V
+15 VD
14.50...15.50 V
+5 V
4.80...5.30 V
VBUS
10.00…17.00 V
+12 V
11.50…12.50 V
+3.3 V
3.00…3.60 V
VBUS
10.00…17.00 V