Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
66
Document no
.
M1144955-001
4.14.1 Sensor
Sensor type:
indicates the type of the sensor
connected.
Lot code:
indicates the manufacturing lot code of the
sensor. The lot code contains the manufacturing
date and shift.
Serial no:
indicates the serial number of the sensor.
Shelf life:
indicates max storage duration.
Usage count:
indicates how many times the sensor
has been attached/detached. Not in use.
Mains Freq.:
indicates the set mains frequency;
50Hz/60 Hz.
Sensor Impedances:
indicates the last measured
impedances.
Imped.
indicates the measured impedance value in
Kohms.
Qualif.
indicates the quality of the measured
impedance; PASS/FAIL.
BE powerup test:
indicates the status of BIS Engine power up test: PASS/FAIL.
DSC selftest ch1:
indicates the DSC selftest status for channel 1: PASS/FAIL.
DSC selftest ch2:
indicates the DSC selftest status for channel 2: PASS/FAIL.
Measured parameters indicated:
BIS
indicates BIS index; range 0..1000 (corresponds to 0..100)
SQI
indicates signal quality index; range 0..1000 (corresponds to 0..100)
EMG
indicates EMG activity level; range 0..10000 (corresponds to 0..100dB)
(NOTE! On the trend display, EMG will be shown between 30..80dB)
SR
indicates supression ratio; range 0...100 (corresponds to 0..100%)
Statuses (HEX):
See
APPENDIX A: How to read HEX numbers
.
DSC status:
indicates the DSC status for the four channels. See
for a detailed description of the message.
Sensor status:
indicates the Sensor status. See
for a
detailed description of the message.
BE powup stat:
indicates the BE power up status. See
for a detailed description of the message.